Back to Search Start Over

Determination of Aluminum in Dialysis Concentrates by Atomic Absorption Spectrometry after Coprecipitation with Lanthanum Phosphate.

Authors :
Selvi EK
Şahin U
Şahan S
Source :
Iranian journal of pharmaceutical research : IJPR [Iran J Pharm Res] 2017 Summer; Vol. 16 (3), pp. 1030-1036.
Publication Year :
2017

Abstract

This method was developed for the determination of trace amounts of aluminum(III) in dialysis concentrates using atomic absorption spectrometry after coprecipitation with lanthanum phosphate. The analytical parameters that influenced the quantitative coprecipitation of analyte including amount of lanthanum, amount of phosfate, pH and duration time were optimized. The % recoveries of the analyte ion were in the range of 95-105 % with limit of detection (3s) of 0.5 µg l <superscript>-1</superscript> . Preconcentration factor was found as 1000 and Relative Standard Deviation (RSD) % value obtained from model solutions was 2.5% for 0.02 mg L <superscript>-1</superscript> . The accuracy of the method was evaluated with standard reference material (CWW-TMD Waste Water). The method was also applied to most concentrated acidic and basic dialysis concentrates with satisfactory results.

Details

Language :
English
ISSN :
1735-0328
Volume :
16
Issue :
3
Database :
MEDLINE
Journal :
Iranian journal of pharmaceutical research : IJPR
Publication Type :
Academic Journal
Accession number :
29201091