Back to Search Start Over

Preparation method and quality control of multigamma volume sources with different matrices.

Authors :
Listkowska A
Lech E
Saganowski P
Tymiński Z
Dziel T
Cacko D
Ziemek T
Kołakowska E
Broda R
Source :
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2018 Apr; Vol. 134, pp. 126-130. Date of Electronic Publication: 2017 Oct 10.
Publication Year :
2018

Abstract

The aim of the work was to develop new radioactive standard sources based on epoxy resins. The optimal proportions of the components and the homogeneity of the matrices were determined. The activity of multigamma sources prepared in Marinelli beakers was determined with reference to the National Standard of Radionuclides Activity in Poland. The difference of radionuclides activity values determined using calibrated gamma spectrometer and the activity of standard solutions used are in most cases significantly lower than measurement uncertainty limits. Sources production method and quality control procedure have been developed.<br /> (Copyright © 2017 Elsevier Ltd. All rights reserved.)

Details

Language :
English
ISSN :
1872-9800
Volume :
134
Database :
MEDLINE
Journal :
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
Publication Type :
Academic Journal
Accession number :
29102162
Full Text :
https://doi.org/10.1016/j.apradiso.2017.10.019