Back to Search Start Over

Low group delay signal conditioning for wearable central blood pressure monitoring device.

Authors :
Fierro G
Silveira F
Armentano R
Source :
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference [Annu Int Conf IEEE Eng Med Biol Soc] 2017 Jul; Vol. 2017, pp. 3285-3288.
Publication Year :
2017

Abstract

Pulse transit time (ptt) is a widely researched approach for wearable, unobtrusive, blood pressure monitoring. The estimation of ptt, being a delay measurement, may be affected by the group delay introduced by the signal conditioning chain. In this work a previously reported method for estimating central (aortic) blood pressure from ptt at aortic domain, using ECG R wave and BCG J wave detection, is considered. A simple design approach for the signal conditioning chain, which is suitable for a wearable device and that takes care of minimizing the impact of eventually introduced spurious delays is presented. The design provides less than 2ms group-delay. The design of a wearable device prototype for ECG, BCG and ptt acquisition and experimental results of its application are reported, showing the effectiveness of the proposed approach.

Details

Language :
English
ISSN :
2694-0604
Volume :
2017
Database :
MEDLINE
Journal :
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Publication Type :
Academic Journal
Accession number :
29060599
Full Text :
https://doi.org/10.1109/EMBC.2017.8037558