Back to Search
Start Over
Low group delay signal conditioning for wearable central blood pressure monitoring device.
- Source :
-
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference [Annu Int Conf IEEE Eng Med Biol Soc] 2017 Jul; Vol. 2017, pp. 3285-3288. - Publication Year :
- 2017
-
Abstract
- Pulse transit time (ptt) is a widely researched approach for wearable, unobtrusive, blood pressure monitoring. The estimation of ptt, being a delay measurement, may be affected by the group delay introduced by the signal conditioning chain. In this work a previously reported method for estimating central (aortic) blood pressure from ptt at aortic domain, using ECG R wave and BCG J wave detection, is considered. A simple design approach for the signal conditioning chain, which is suitable for a wearable device and that takes care of minimizing the impact of eventually introduced spurious delays is presented. The design provides less than 2ms group-delay. The design of a wearable device prototype for ECG, BCG and ptt acquisition and experimental results of its application are reported, showing the effectiveness of the proposed approach.
Details
- Language :
- English
- ISSN :
- 2694-0604
- Volume :
- 2017
- Database :
- MEDLINE
- Journal :
- Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
- Publication Type :
- Academic Journal
- Accession number :
- 29060599
- Full Text :
- https://doi.org/10.1109/EMBC.2017.8037558