Measurement of water vapor using an imaging field-widened spatial heterodyne spectrometer.

Authors :
Langille JA
Solheim B
Bourassa A
Degenstein D
Brown S
Shepherd GG
Source :
Applied optics [Appl Opt] 2017 May 20; Vol. 56 (15), pp. 4297-4308.
Publication Year :
2017

Abstract

Spatial heterodyne spectroscopy (SHS) has been successfully applied to measure variations in water vapor pressure in the lab. An imaging system is combined with a monolithic field-widened SHS to observe a white-light source through a 1 m length water vapor cell that is designed to produce predictable variations in the water vapor pressure. The performance of the spectrometer design is examined by comparing spectra simulated using a radiative transfer model to observed spectra at several cell pressures. The intended application of the instrument to vertically resolve the water vapor profile in the upper troposphere and lower stratosphere using limb-scattered radiation in a vibrational band of water (1363-1366 nm) is also introduced.