Measurements of air kerma index in computed tomography: A comparison among methodologies.
- Source :
-
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2018 Aug; Vol. 138, pp. 10-13. Date of Electronic Publication: 2017 Oct 05. - Publication Year :
- 2018
-
Abstract
- As CT exams impart high doses to patients in comparison to other radiologist techniques, reliable dosimetry is required. In this work, dosimetry in CT beams was carried out in terms of air kerma index in air or in a phantom measured by a pencil ionization chamber, thermoluminescent (TL) detectors and radiochromic film. Calibration results showed the low energy dependence of all three dosimetric systems for the 100-120kV range, the very high uncertainty of the TL dosimeters in comparison to the other systems and high nonlinearity response in terms of air kerma of the radiochromic film. Measurements with the three systems in a 120kV CT protocol showed an acceptable agreement among the weighted air kerma index values, but TL dosimeters presented the highest uncertainties in the values.<br /> (Copyright © 2017 Elsevier Ltd. All rights reserved.)
- Subjects :
- Film Dosimetry methods
Film Dosimetry statistics & numerical data
Humans
Phantoms, Imaging
Radiometry instrumentation
Radiometry statistics & numerical data
Thermoluminescent Dosimetry methods
Thermoluminescent Dosimetry statistics & numerical data
X-Ray Film
Radiometry methods
Tomography, X-Ray Computed
Subjects
Details
- Language :
- English
- ISSN :
- 1872-9800
- Volume :
- 138
- Database :
- MEDLINE
- Journal :
- Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
- Publication Type :
- Academic Journal
- Accession number :
- 29033210
- Full Text :
- https://doi.org/10.1016/j.apradiso.2017.10.008