Measurements of air kerma index in computed tomography: A comparison among methodologies.

Authors :
Alonso TC
Mourão Filho AP
Da Silva TA
Source :
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2018 Aug; Vol. 138, pp. 10-13. Date of Electronic Publication: 2017 Oct 05.
Publication Year :
2018

Abstract

As CT exams impart high doses to patients in comparison to other radiologist techniques, reliable dosimetry is required. In this work, dosimetry in CT beams was carried out in terms of air kerma index in air or in a phantom measured by a pencil ionization chamber, thermoluminescent (TL) detectors and radiochromic film. Calibration results showed the low energy dependence of all three dosimetric systems for the 100-120kV range, the very high uncertainty of the TL dosimeters in comparison to the other systems and high nonlinearity response in terms of air kerma of the radiochromic film. Measurements with the three systems in a 120kV CT protocol showed an acceptable agreement among the weighted air kerma index values, but TL dosimeters presented the highest uncertainties in the values.<br /> (Copyright © 2017 Elsevier Ltd. All rights reserved.)

Details

Language :
English
ISSN :
1872-9800
Volume :
138
Database :
MEDLINE
Journal :
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
Publication Type :
Academic Journal
Accession number :
29033210
Full Text :
https://doi.org/10.1016/j.apradiso.2017.10.008