Back to Search
Start Over
Design and Fabrication of MR-Tracked Metallic Stylet for Gynecologic Brachytherapy.
- Source :
-
IEEE/ASME transactions on mechatronics : a joint publication of the IEEE Industrial Electronics Society and the ASME Dynamic Systems and Control Division [IEEE ASME Trans Mechatron] 2016 Apr; Vol. 21 (2), pp. 956-962. Date of Electronic Publication: 2015 Nov 24. - Publication Year :
- 2016
-
Abstract
- Active magnetic resonance (MR) tracking for gynecologic brachytherapy was made possible by attaching the micro radiofrequency coils to the brachytherapy applicator. The rectangular planar micro coil was fabricated using flexible printed circuits with dimensions of 8mm×1.5mm. A 5-Fr (1.6mm) tungsten brachytherapy stylet was custom-machined to incorporate the micro coils. The finite element analysis and the phantom tissue studies show that the proposed device enables in situ , real-time guidance of access routes to the target anatomy safely and accurately. The setup was tested in a Siemens 3T MR scanner. The micro coils can be localized rapidly (up to 40 Hz) and precisely (resolution: 0.6×0.6×0.6mm <superscript>3</superscript> ) using an MR-tracking sequence.
Details
- Language :
- English
- ISSN :
- 1083-4435
- Volume :
- 21
- Issue :
- 2
- Database :
- MEDLINE
- Journal :
- IEEE/ASME transactions on mechatronics : a joint publication of the IEEE Industrial Electronics Society and the ASME Dynamic Systems and Control Division
- Publication Type :
- Academic Journal
- Accession number :
- 28989272
- Full Text :
- https://doi.org/10.1109/TMECH.2015.2503427