Back to Search Start Over

Comparative analysis of condom lubricants on pre & post-coital vaginal swabs using AccuTOF-DART.

Authors :
Proni G
Cohen P
Huggins LA
Nesnas N
Source :
Forensic science international [Forensic Sci Int] 2017 Nov; Vol. 280, pp. 87-94. Date of Electronic Publication: 2017 Sep 28.
Publication Year :
2017

Abstract

In this study we demonstrate the use of Direct Analysis in Real Time Mass Spectrometry (DART) as a powerful tool for detection of nonoxynol in vaginal fluid post contact with a condom, enabling rapid tracing and added evidences in sexual assault crimes. Vaginal fluid was sampled using cotton swabs and glass rods and measured directly with DART. Sample preparation using water, hexane, methanol, and dichloromethane extraction, was explored for comparison and optimization of signals. Nonoxynol was detected up to eight hours after sampling. Optimal sampling conditions and mass spectrometry parameters are reported and discussed.<br /> (Copyright © 2017 Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1872-6283
Volume :
280
Database :
MEDLINE
Journal :
Forensic science international
Publication Type :
Academic Journal
Accession number :
28965000
Full Text :
https://doi.org/10.1016/j.forsciint.2017.09.005