Back to Search Start Over

Validation of the Omron HEM-7320-LA, upper arm blood pressure monitor with Intelli Wrap Technology Cuff HEM-FL1 for self-measurement and clinic use according to the European Society of Hypertension International Protocol revision 2010 in the Mexican population.

Authors :
Grover-Páez F
Cardona-Muñoz EG
Cardona-Müller D
Guzmán-Saldívar VH
Rodríguez-De la Cerda M
Jiménez-Cázarez MB
Totsuka-Sutto SE
Alanis-Sánchez GA
Ramos-Becerra CG
Source :
Blood pressure monitoring [Blood Press Monit] 2017 Dec; Vol. 22 (6), pp. 375-378.
Publication Year :
2017

Abstract

Objective: The aim of this study was to determine the accuracy of the Omron HEM-7320-LA with Intelli Wrap technology cuff HEM-FL1 for self-measurement and clinic blood pressure (BP) measurement according to the European Society of Hypertension International Protocol revision 2010.<br />Participants and Methods: The evaluation was performed in 39 individuals. The mean age of the participants was 47.9±14 years; systolic BP was 145.2±24.3 mmHg (range: 97-190), diastolic BP was 90.9±12.9 mmHg (range: 68-120), and arm circumference was 30.8±4 cm (range: 25-38.5).<br />Results: The device successfully fulfilled the established criteria of the validation protocol. The device overestimated systolic BP by 0.6±5.7 mmHg and diastolic BP by 2.2±5.1 mmHg. The specially designed cuff HEM-FL1 to cover a broad range of arm circumferences and self-placement fulfilled the requirements of the International Protocol.

Details

Language :
English
ISSN :
1473-5725
Volume :
22
Issue :
6
Database :
MEDLINE
Journal :
Blood pressure monitoring
Publication Type :
Academic Journal
Accession number :
28945665
Full Text :
https://doi.org/10.1097/MBP.0000000000000290