Determination of 74 new psychoactive substances in serum using automated in-line solid-phase extraction-liquid chromatography-tandem mass spectrometry.

Authors :
Lehmann S
Kieliba T
Beike J
Thevis M
Mercer-Chalmers-Bender K
Source :
Journal of chromatography. B, Analytical technologies in the biomedical and life sciences [J Chromatogr B Analyt Technol Biomed Life Sci] 2017 Oct 01; Vol. 1064, pp. 124-138. Date of Electronic Publication: 2017 Sep 07.
Publication Year :
2017

Abstract

A detailed description is given of the development and validation of a fully automated in-line solid-phase extraction-liquid chromatography-tandem mass spectrometry (SPE-LC-MS/MS) method capable of detecting 90 central-stimulating new psychoactive substances (NPS) and 5 conventional amphetamine-type stimulants (amphetamine, 3,4-methylenedioxy-methamphetamine (MDMA), 3,4-methylenedioxy-amphetamine (MDA), 3,4-methylenedioxy-N-ethyl-amphetamine (MDEA), methamphetamine) in serum. The aim was to apply the validated method to forensic samples. The preparation of 150μL of serum was performed by an Instrument Top Sample Preparation (ITSP)-SPE with mixed mode cation exchanger cartridges. The extracts were directly injected into an LC-MS/MS system, using a biphenyl column and gradient elution with 2mM ammonium formate/0.1% formic acid and acetonitrile/0.1% formic acid as mobile phases. The chromatographic run time amounts to 9.3min (including re-equilibration). The total cycle time is 11min, due to the interlacing between sample preparation and analysis. The method was fully validated using 69 NPS and five conventional amphetamine-type stimulants, according to the guidelines of the Society of Toxicological and Forensic Chemistry (GTFCh). The guidelines were fully achieved for 62 analytes (with a limit of detection (LOD) between 0.2 and 4μg/L), whilst full validation was not feasible for the remaining 12 analytes. For the fully validated analytes, the method achieved linearity in the 5μg/L (lower limit of quantification, LLOQ) to 250μg/L range (coefficients of determination>0.99). Recoveries for 69 of these compounds were greater than 50%, with relative standard deviations≤15%. The validated method was then tested for its capability in detecting a further 21 NPS, thus totalling 95 tested substances. An LOD between 0.4 and 1.6μg/L was obtained for these 21 additional qualitatively-measured substances. The method was subsequently successfully applied to 28 specimens from routine forensic case work, of which 7 samples were determined to be positive for NPS consumption.<br /> (Copyright © 2017 Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1873-376X
Volume :
1064
Database :
MEDLINE
Journal :
Journal of chromatography. B, Analytical technologies in the biomedical and life sciences
Publication Type :
Academic Journal
Accession number :
28922649
Full Text :
https://doi.org/10.1016/j.jchromb.2017.09.003