Back to Search
Start Over
Passive Microlesion Detection and Mapping for Treatment of Hypertrophic Cardiomyopathy.
- Source :
-
AIP conference proceedings [AIP Conf Proc] 2017; Vol. 1816 (1). Date of Electronic Publication: 2017 Mar 01. - Publication Year :
- 2017
-
Abstract
- Intermittent high intensity ultrasound pulses with circulating contrast agent microbubbles can induce scattered microlesions of potential value for myocardial reduction therapy. This paper presents an in vitro setup imitating the treatment for monitoring development. A preclinical imaging system with a single element transducer, synchronization and receive-only imaging transducer array has been implemented on a research platform. Contrast agent microbubbles pumped in a dialysis tubing setup were exposed to high intensity focused ultrasound at 1.0/3.5 MHz center frequencies. Polystyrene spheres were employed as linear scatterers compared to contrast agents for system transfer function equalization. A cavitation mapping technique was employed to spatially locate and depict microbubble activity during treatment. For high acoustic pressure amplitudes a 5 dB difference between contrast agent and solid spheres was observed and spatially mapped. The in-plane resolution was 4.5 mm for axial and 1.5 mm laterally. In the future, this cavitation detection scheme will be applied to monitor in vivo microlesioning in real-time.
Details
- Language :
- English
- ISSN :
- 0094-243X
- Volume :
- 1816
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- AIP conference proceedings
- Publication Type :
- Academic Journal
- Accession number :
- 28757665
- Full Text :
- https://doi.org/10.1063/1.4976594