Acute Structural Failure of the Trifecta Aortic Valve Bioprosthesis.

Authors :
Zhu MZL
Newman MA
Joshi P
Passage J
Source :
Heart, lung & circulation [Heart Lung Circ] 2017 Nov; Vol. 26 (11), pp. e82-e85. Date of Electronic Publication: 2017 Apr 20.
Publication Year :
2017

Abstract

Advancements in bioprosthetic valve technology have greatly enhanced the haemodynamic performance and long-term durability of tissue valves. These features, along with the key advantage of avoiding lifelong anticoagulation, have made bioprosthetic valves increasingly attractive for clinicians and patients alike. The St Jude Trifecta valve is a novel, bovine pericardial bioprosthesis with promising early data for performance and safety. However, no prosthetic valve is perfect and prosthesis failure can occur with even the most reliable and well-designed devices. We report a case of early and acute structural deterioration (stent-post leaflet rupture) of the Trifecta valve, explanted after 33 months, in a 76-year-old male.<br /> (Crown Copyright © 2017. Published by Elsevier B.V. All rights reserved.)