Back to Search
Start Over
In vitro and in vivo evaluation of a novel integrated wearable artificial lung.
- Source :
-
The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation [J Heart Lung Transplant] 2017 Jul; Vol. 36 (7), pp. 806-811. Date of Electronic Publication: 2017 Mar 04. - Publication Year :
- 2017
-
Abstract
- Background: Conventional extracorporeal membrane oxygenation (ECMO) is cumbersome and is associated with high morbidity and mortality. We are currently developing the Pittsburgh Ambulatory Assist Lung (PAAL), which is designed to allow for ambulation of lung failure patients during bridge to transplant or recovery. In this study, we investigated the in vitro and acute in vivo performance of the PAAL.<br />Methods: The PAAL features a 1.75-inch-diameter, cylindrical, hollow-fiber membrane (HFM) bundle of stacked sheets, with a surface area of 0.65 m <superscript>2</superscript> integrated with a centrifugal pump. The PAAL was tested on the bench for hydrodynamic performance, gas exchange and hemolysis. It was then tested in 40- to 60-kg adult sheep (n = 4) for 6 hours. The animals were cannulated with an Avalon Elite 27Fr dual-lumen catheter (DLC) inserted through the right external jugular into the superior vena cava (SVC), right atrium (RA) and inferior vena cava (IVC).<br />Results: The PAAL pumped >250 mm Hg at 3.5 liters/min at a rotation speed of 2,100 rpm. Oxygenation performance met the target of 180 ml/min at 3.5 liters/min of blood flow in vitro, resulting in a gas-exchange efficiency of 278 ml/min/m <superscript>2</superscript> . The normalized index of hemolysis (NIH) for the PAAL and cannula was 0.054 g per 100 liters (n = 2) at 3.5 liters/min, as compared with 0.020 g per 100 liters (n = 2) for controls (DLC cannula and a Centrimag pump). Plasma-free hemoglobin (pfHb) was <20 mg/dl for all animals. Blood left the device 100% oxygenated in vivo and oxygenation reached 181 ml/min at 3.8 liters/min.<br />Conclusion: The PAAL met in vitro and acute in vivo performance targets. Five-day chronic sheep studies are planned for the near future.<br /> (Copyright © 2017 International Society for the Heart and Lung Transplantation. Published by Elsevier Inc. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1557-3117
- Volume :
- 36
- Issue :
- 7
- Database :
- MEDLINE
- Journal :
- The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation
- Publication Type :
- Academic Journal
- Accession number :
- 28359655
- Full Text :
- https://doi.org/10.1016/j.healun.2017.02.025