Back to Search Start Over

Analysis of Monoclonal Antibodies in Human Serum as a Model for Clinical Monoclonal Gammopathy by Use of 21 Tesla FT-ICR Top-Down and Middle-Down MS/MS.

Authors :
He L
Anderson LC
Barnidge DR
Murray DL
Hendrickson CL
Marshall AG
Source :
Journal of the American Society for Mass Spectrometry [J Am Soc Mass Spectrom] 2017 May; Vol. 28 (5), pp. 827-838. Date of Electronic Publication: 2017 Feb 28.
Publication Year :
2017

Abstract

With the rapid growth of therapeutic monoclonal antibodies (mAbs), stringent quality control is needed to ensure clinical safety and efficacy. Monoclonal antibody primary sequence and post-translational modifications (PTM) are conventionally analyzed with labor-intensive, bottom-up tandem mass spectrometry (MS/MS), which is limited by incomplete peptide sequence coverage and introduction of artifacts during the lengthy analysis procedure. Here, we describe top-down and middle-down approaches with the advantages of fast sample preparation with minimal artifacts, ultrahigh mass accuracy, and extensive residue cleavages by use of 21 tesla FT-ICR MS/MS. The ultrahigh mass accuracy yields an RMS error of 0.2-0.4 ppm for antibody light chain, heavy chain, heavy chain Fc/2, and Fd subunits. The corresponding sequence coverages are 81%, 38%, 72%, and 65% with MS/MS RMS error ~4 ppm. Extension to a monoclonal antibody in human serum as a monoclonal gammopathy model yielded 53% sequence coverage from two nano-LC MS/MS runs. A blind analysis of five therapeutic monoclonal antibodies at clinically relevant concentrations in human serum resulted in correct identification of all five antibodies. Nano-LC 21 T FT-ICR MS/MS provides nonpareil mass resolution, mass accuracy, and sequence coverage for mAbs, and sets a benchmark for MS/MS analysis of multiple mAbs in serum. This is the first time that extensive cleavages for both variable and constant regions have been achieved for mAbs in a human serum background. Graphical Abstract ᅟ.

Details

Language :
English
ISSN :
1879-1123
Volume :
28
Issue :
5
Database :
MEDLINE
Journal :
Journal of the American Society for Mass Spectrometry
Publication Type :
Academic Journal
Accession number :
28247297
Full Text :
https://doi.org/10.1007/s13361-017-1602-6