Back to Search Start Over

Development of a method for rapid analysis of Ra-226 in groundwater and discharge water samples by ICP-QQQ-MS.

Authors :
van Es EM
Russell BC
Ivanov P
Read D
Source :
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2017 Aug; Vol. 126, pp. 31-34. Date of Electronic Publication: 2017 Feb 20.
Publication Year :
2017

Abstract

A novel and rapid method has been developed for pre-concentration and measurement of <superscript>226</superscript> Ra in groundwater and discharge water samples using the latest generation triple quadrupole inductively coupled plasma mass spectrometry (ICP-QQQ-MS). Cation exchange and extraction chromatography are capable of pre-concentration factors of ~200 based on 1L samples. The sensitivity and interference removal capability of ICP-QQQ-MS was assessed from spiked groundwaters, with the introduction of He collision gas required to minimise instrument background in high-matrix samples. The technique developed is potentially capable of detecting <superscript>226</superscript> Ra activities as low as 5mBqL <superscript>-1</superscript> when combined with pre-concentration prior to measurement.<br /> (Crown Copyright © 2017. Published by Elsevier Ltd. All rights reserved.)

Details

Language :
English
ISSN :
1872-9800
Volume :
126
Database :
MEDLINE
Journal :
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
Publication Type :
Academic Journal
Accession number :
28236556
Full Text :