Characterization of micro-resonator based on enhanced metal insulator semiconductor capacitor for glucose recognition.
- Source :
-
Medical engineering & physics [Med Eng Phys] 2017 Mar; Vol. 41, pp. 55-62. Date of Electronic Publication: 2017 Jan 31. - Publication Year :
- 2017
-
Abstract
- We present a concept for the characterization of micro-fabricated based resonator incorporating air-bridge metal-insulator-semiconductor (MIS) capacitor to continuously monitor an individual's state of glucose levels based on frequency variation. The investigation revealed that, the micro-resonator based on MIS capacitor holds considerable promise for implementation and recognition as a glucose sensor for human serum. The discrepancy in complex permittivity as a result of enhanced capacitor was achieved for the detection and determination of random glucose concentration levels using a unique variation of capacitor that indeed results in an adequate variation of the resonance frequency. Moreover, the design and development of micro-resonator with enhanced MIS capacitor generate a resolution of 112.38 × 10 <superscript>-3</superscript> pF/mg/dl, minimum detectable glucose level of 7.45mg/dl, and a limit of quantification of 22.58mg/dl. Additionally, this unique approach offers long-term reliability for mediator-free glucose sensing with a relative standard deviation of less than 0.5%.<br /> (Copyright © 2017 IPEM. Published by Elsevier Ltd. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1873-4030
- Volume :
- 41
- Database :
- MEDLINE
- Journal :
- Medical engineering & physics
- Publication Type :
- Academic Journal
- Accession number :
- 28159448
- Full Text :
- https://doi.org/10.1016/j.medengphy.2017.01.008