Back to Search Start Over

Study of Optimal Perimetric Testing In Children (OPTIC): development and feasibility of the kinetic perimetry reliability measure (KPRM).

Authors :
Patel DE
Viswanathan AC
Garway-Heath D
Cumberland PM
Walters BC
Russell-Eggitt I
Cortina-Borja M
Rahi JS
Source :
The British journal of ophthalmology [Br J Ophthalmol] 2017 Feb; Vol. 101 (2), pp. 94-96. Date of Electronic Publication: 2016 Nov 08.
Publication Year :
2017

Abstract

Introduction: Interpretation of perimetric findings, particularly in children, relies on accurate assessment of test reliability, yet no objective measures of reliability exist for kinetic perimetry. We developed the kinetic perimetry reliability measure (KPRM), a quantitative measure of perimetric test reproducibility/reliability and report here its feasibility and association with subjective assessment of reliability.<br />Methods: Children aged 5-15 years, without an ophthalmic condition that affects the visual field, were recruited from Moorfields Eye Hospital and underwent Goldmann perimetry as part of a wider research programme on perimetry in children. Subjects were tested with two isopters and the blind spot was plotted, followed by a KPRM. Test reliability was also scored qualitatively using our examiner-based assessment of reliability (EBAR) scoring system, which standardises the conventional clinical approach to assessing test quality. The relationship between KPRM and EBAR was examined to explore the use of KPRM in assessing reliability of kinetic fields.<br />Results: A total of 103 children (median age 8.9 years; IQR: 7.1 to 11.8 years) underwent Goldmann perimetry with KPRM and EBAR scoring. A KPRM was achieved by all children. KPRM values increased with reducing test quality (Kruskal-Wallis, p=0.005), indicating greater test-retest variability, and reduced with age (linear regression, p=0.015). One of 103 children (0.97%) demonstrated discordance between EBAR and KPRM.<br />Conclusion: KPRM and EBAR are distinct but complementary approaches. Though scores show excellent agreement, KPRM is able to quantify within-test variability, providing data not captured by subjective assessment. Thus, we suggest combining KPRM with EBAR to aid interpretation of kinetic perimetry test reliability in children.<br /> (Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://www.bmj.com/company/products-services/rights-and-licensing/.)

Details

Language :
English
ISSN :
1468-2079
Volume :
101
Issue :
2
Database :
MEDLINE
Journal :
The British journal of ophthalmology
Publication Type :
Academic Journal
Accession number :
28108479
Full Text :
https://doi.org/10.1136/bjophthalmol-2016-309402