Evaluation of artifacts generated by zirconium implants in cone-beam computed tomography images.

Authors :
Vasconcelos TV
Bechara BB
McMahan CA
Freitas DQ
Noujeim M
Source :
Oral surgery, oral medicine, oral pathology and oral radiology [Oral Surg Oral Med Oral Pathol Oral Radiol] 2017 Feb; Vol. 123 (2), pp. 265-272. Date of Electronic Publication: 2016 Nov 01.
Publication Year :
2017

Abstract

Objective: To evaluate zirconium implant artifact production in cone beam computed tomography images obtained with different protocols.<br />Study Design: One zirconium implant was inserted in an edentulous mandible. Twenty scans were acquired with a ProMax 3D unit (Planmeca Oy, Helsinki, Finland), with acquisition settings ranging from 70 to 90 peak kilovoltage (kVp) and voxel sizes of 0.32 and 0.16 mm. A metal artifact reduction (MAR) tool was activated in half of the scans. An axial slice through the middle region of the implant was selected for each dataset. Gray values (mean ± standard deviation) were measured in two regions of interest, one close to and the other distant from the implant (control area). The contrast-to-noise ratio was also calculated.<br />Results: Standard deviation decreased with greater kVp and when the MAR tool was used. The contrast-to-noise ratio was significantly higher when the MAR tool was turned off, except for low resolution with kVp values above 80. Selection of the MAR tool and greater kVp resulted in an overall reduction of artifacts in images acquired with low resolution.<br />Conclusions: Although zirconium implants do produce image artifacts in cone-bean computed tomography scans, the setting that best controlled artifact generation by zirconium implants was 90 kVp at low resolution and with the MAR tool turned on.<br /> (Copyright © 2016 Elsevier Inc. All rights reserved.)

Details

Language :
English
ISSN :
2212-4411
Volume :
123
Issue :
2
Database :
MEDLINE
Journal :
Oral surgery, oral medicine, oral pathology and oral radiology
Publication Type :
Academic Journal
Accession number :
28086998
Full Text :
https://doi.org/10.1016/j.oooo.2016.10.021