A new method to calibrate the absolute sensitivity of a soft X-ray streak camera.

Authors :
Yu J
Liu S
Li J
Yang Z
Chen M
Guo L
Yao L
Xiao S
Source :
The Review of scientific instruments [Rev Sci Instrum] 2016 Dec; Vol. 87 (12), pp. 123506.
Publication Year :
2016

Abstract

In this paper, we introduce a new method to calibrate the absolute sensitivity of a soft X-ray streak camera (SXRSC). The calibrations are done in the static mode by using a small laser-produced X-ray source. A calibrated X-ray CCD is used as a secondary standard detector to monitor the X-ray source intensity. In addition, two sets of holographic flat-field grating spectrometers are chosen as the spectral discrimination systems of the SXRSC and the X-ray CCD. The absolute sensitivity of the SXRSC is obtained by comparing the signal counts of the SXRSC to the output counts of the X-ray CCD. Results show that the calibrated spectrum covers the range from 200 eV to 1040 eV. The change of the absolute sensitivity in the vicinity of the K-edge of the carbon can also be clearly seen. The experimental values agree with the calculated values to within 29% error. Compared with previous calibration methods, the proposed method has several advantages: a wide spectral range, high accuracy, and simple data processing. Our calibration results can be used to make quantitative X-ray flux measurements in laser fusion research.

Details

Language :
English
ISSN :
1089-7623
Volume :
87
Issue :
12
Database :
MEDLINE
Journal :
The Review of scientific instruments
Publication Type :
Academic Journal
Accession number :
28040978
Full Text :
https://doi.org/10.1063/1.4971847