Back to Search
Start Over
A streaked X-ray spectroscopy platform for rapidly heated, near-solid density plasmas.
- Source :
-
The Review of scientific instruments [Rev Sci Instrum] 2016 Nov; Vol. 87 (11), pp. 11E312. - Publication Year :
- 2016
-
Abstract
- A picosecond, time-resolved, x-ray spectroscopy platform was developed to study the thermal line emission from rapidly heated solid targets containing buried aluminum or iron layers. The targets were driven by high-contrast 1ω or 2ω laser pulses at focused intensities up to 1 × 10 <superscript>19</superscript> W/cm <superscript>2</superscript> . The experimental platform combines time-integrating and time-resolved x-ray spectrometers. Picosecond time resolution was achieved with a pair of ultrafast x-ray streak cameras coupled to high-throughput Hall spectrometers. Time-integrated spectra were collected on each shot to correct the streaked data for variations in x-ray photocathode spectral sensitivity. The time-integrated spectrometer uses three elliptical crystals to disperse x rays with energies between 800 and 2100 eV with moderate (E/ΔE ∼ 450) resolving power. The streaked spectrometers accept four interchangeable conical crystals with higher resolving power (E/ΔE ∼ 650) to measure the brightest thermal lines in the 1300 to 1700 eV spectral range.
Details
- Language :
- English
- ISSN :
- 1089-7623
- Volume :
- 87
- Issue :
- 11
- Database :
- MEDLINE
- Journal :
- The Review of scientific instruments
- Publication Type :
- Academic Journal
- Accession number :
- 27910646
- Full Text :
- https://doi.org/10.1063/1.4960414