A prototype of a picosecond x-ray streak camera has been developed and tested by Commissariat à l'Énergie Atomique et aux Énergies Alternatives to provide plasma-diagnostic support for the Laser Megajoule. We report on the measured performance of this streak camera, which almost fulfills the requirements: 50-μm spatial resolution over a 15-mm field in the photocathode plane, 17-ps temporal resolution in a 2-ns timebase, a detection threshold lower than 625 nJ/cm <superscript>2</superscript> in the 0.05-15 keV spectral range, and a dynamic range greater than 100.