Back to Search
Start Over
Performance of Laser Megajoule's x-ray streak camera.
- Source :
-
The Review of scientific instruments [Rev Sci Instrum] 2016 Nov; Vol. 87 (11), pp. 11E303. - Publication Year :
- 2016
-
Abstract
- A prototype of a picosecond x-ray streak camera has been developed and tested by Commissariat à l'Énergie Atomique et aux Énergies Alternatives to provide plasma-diagnostic support for the Laser Megajoule. We report on the measured performance of this streak camera, which almost fulfills the requirements: 50-μm spatial resolution over a 15-mm field in the photocathode plane, 17-ps temporal resolution in a 2-ns timebase, a detection threshold lower than 625 nJ/cm <superscript>2</superscript> in the 0.05-15 keV spectral range, and a dynamic range greater than 100.
Details
- Language :
- English
- ISSN :
- 1089-7623
- Volume :
- 87
- Issue :
- 11
- Database :
- MEDLINE
- Journal :
- The Review of scientific instruments
- Publication Type :
- Academic Journal
- Accession number :
- 27910354
- Full Text :
- https://doi.org/10.1063/1.4959165