Accuracy of pulse interval timing in ambulatory blood pressure measurement.

Authors :
Kane SA
Blake JR
McArdle FJ
Langley P
Sims AJ
Source :
Scientific reports [Sci Rep] 2016 Nov 23; Vol. 6, pp. 37472. Date of Electronic Publication: 2016 Nov 23.
Publication Year :
2016

Abstract

Blood pressure (BP) monitors rely on pulse detection. Some blood pressure monitors use pulse timings to analyse pulse interval variability for arrhythmia screening, but this assumes that the pulse interval timings detected from BP cuffs are accurate compared with RR intervals derived from ECG. In this study we compared the accuracy of pulse intervals detected using an ambulatory blood pressure monitor (ABPM) with single lead ECG. Twenty participants wore an ABPM for three hours and a data logger which synchronously measured cuff pressure and ECG. RR intervals were compared with corresponding intervals derived from the cuff pressure tracings using three different pulse landmarks. Linear mixed effects models were used to assess differences between ECG and cuff pressure timings and to investigate the effect of potential covariates. In addition, the maximum number of successive oscillometric beats detectable in a measurement was assessed. From 243 BP measurements, the landmark at the foot of the oscillometric pulse was found to be associated with fewest covariates and had a random error of 9.5 ms. 99% of the cuff pressure recordings had more than 10 successive detectable oscillometric beats. RR intervals can be accurately estimated using an ABPM.

Details

Language :
English
ISSN :
2045-2322
Volume :
6
Database :
MEDLINE
Journal :
Scientific reports
Publication Type :
Academic Journal
Accession number :
27876841
Full Text :
https://doi.org/10.1038/srep37472