Back to Search Start Over

Voltammetric Detection of Urea on an Ag-Modified Zeolite-Expanded Graphite-Epoxy Composite Electrode.

Authors :
Manea F
Pop A
Radovan C
Malchev P
Bebeselea A
Burtica G
Picken S
Schoonman J
Source :
Sensors (Basel, Switzerland) [Sensors (Basel)] 2008 Sep 22; Vol. 8 (9), pp. 5806-5819. Date of Electronic Publication: 2008 Sep 22.
Publication Year :
2008

Abstract

In this paper, a modified expanded graphite composite electrode based on natural zeolitic volcanic tuff modified with silver (EG-Ag-Z-Epoxy) was developed. Cyclic voltammetry measurements revealed a reasonably fast electron transfer and a good stability of the electrode in 0.1 M NaOH supporting electrolyte. This modified electrode exhibited moderate electrocatalytic effect towards urea oxidation, allowing its determination in aqueous solution. The linear dependence of the current versus urea concentration was reached using square-wave voltammetry in the concentrations range of urea between 0.2 to 1.4 mM, with a relatively low limit of detection of 0.05 mM. A moderate enhancement of electroanalytical sensitivity for the determination of urea at EG-Ag-Z-Epoxy electrode was reached by applying a chemical preconcentration step prior to voltammetric/amperometric quantification.

Details

Language :
English
ISSN :
1424-8220
Volume :
8
Issue :
9
Database :
MEDLINE
Journal :
Sensors (Basel, Switzerland)
Publication Type :
Academic Journal
Accession number :
27873841
Full Text :
https://doi.org/10.3390/s8095806