Impact of Real-Time Image Gating on Spot Scanning Proton Therapy for Lung Tumors: A Simulation Study.
- Source :
-
International journal of radiation oncology, biology, physics [Int J Radiat Oncol Biol Phys] 2017 Jan 01; Vol. 97 (1), pp. 173-181. Date of Electronic Publication: 2016 Sep 28. - Publication Year :
- 2017
-
Abstract
- Purpose: To investigate the effectiveness of real-time-image gated proton beam therapy for lung tumors and to establish a suitable size for the gating window (GW).<br />Methods and Materials: A proton beam gated by a fiducial marker entering a preassigned GW (as monitored by 2 fluoroscopy units) was used with 7 lung cancer patients. Seven treatment plans were generated: real-time-image gated proton beam therapy with GW sizes of ±1, 2, 3, 4, 5, and 8 mm and free-breathing proton therapy. The prescribed dose was 70 Gy (relative biological effectiveness)/10 fractions to 99% of the target. Each of the 3-dimensional marker positions in the time series was associated with the appropriate 4-dimensional computed tomography phase. The 4-dimensional dose calculations were performed. The dose distribution in each respiratory phase was deformed into the end-exhale computed tomography image. The D99 and D5 to D95 of the clinical target volume scaled by the prescribed dose with criteria of D99 >95% and D5 to D95 <5%, V20 for the normal lung, and treatment times were evaluated.<br />Results: Gating windows ≤ ±2 mm fulfilled the CTV criteria for all patients (whereas the criteria were not always met for GWs ≥ ±3 mm) and gave an average reduction in V20 of more than 17.2% relative to free-breathing proton therapy (whereas GWs ≥ ±4 mm resulted in similar or increased V20). The average (maximum) irradiation times were 384 seconds (818 seconds) for the ±1-mm GW, but less than 226 seconds (292 seconds) for the ±2-mm GW. The maximum increased considerably at ±1-mm GW.<br />Conclusion: Real-time-image gated proton beam therapy with a GW of ±2 mm was demonstrated to be suitable, providing good dose distribution without greatly extending treatment time.<br /> (Copyright © 2016 Elsevier Inc. All rights reserved.)
- Subjects :
- Carcinoma, Non-Small-Cell Lung diagnostic imaging
Carcinoma, Non-Small-Cell Lung pathology
Four-Dimensional Computed Tomography methods
Humans
Lung Neoplasms pathology
Movement
Proton Therapy instrumentation
Radiotherapy Dosage
Respiration
Simulation Training methods
Time Factors
Carcinoma, Non-Small-Cell Lung radiotherapy
Fiducial Markers
Lung Neoplasms diagnostic imaging
Lung Neoplasms radiotherapy
Proton Therapy methods
Radiotherapy Planning, Computer-Assisted methods
Subjects
Details
- Language :
- English
- ISSN :
- 1879-355X
- Volume :
- 97
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- International journal of radiation oncology, biology, physics
- Publication Type :
- Academic Journal
- Accession number :
- 27856039
- Full Text :
- https://doi.org/10.1016/j.ijrobp.2016.09.027