Abstract
- Hemoglobin A1c (HbA1c) is a frequently requested laboratory test and there is thus a need for high throughput instruments for this assay. We evaluated a new automated multicapillary zone electrophoresis instrument (Capillarys 3 Tera, Sebia, Lisses, France) for analysis of HbA1c in venous samples. Routine requested HbA1c samples were analyzed immunologically on a Roche c6000 instrument (n = 142) and then with the Capillarys 3 Tera instrument. The Capillarys 3 Tera instrument performed approximately 70 HbA1c tests/hour. There was a strong linear correlation between Capillarys 3 Tera and Roche Tina-Quant HbA1c Gen 3 assay (y = 1.003x - 0.3246 R <superscript>2 </superscript> = .996). The total CV for the 12 capillaries varied between 0.8 and 2.2% and there was a good agreement between duplicate samples (R <superscript>2 </superscript> = .997). In conclusion, the Capillarys 3 Tera instrument has a high assay capacity for HbA1c. It has a good precision and agreement with the Roche Tina-Quant HbA1c method and is well suited for high volume testing of HbA1c.
- Subjects :
- Automation, Laboratory instrumentation
Electrophoresis, Capillary instrumentation
Hemoglobinometry instrumentation
Humans
Immunoassay instrumentation
Reproducibility of Results
Sensitivity and Specificity
Electrophoresis, Capillary standards
Glycated Hemoglobin analysis
Hemoglobinometry standards
Subjects
Details
- Language :
- English
- ISSN :
- 1502-7686
- Volume :
- 77
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Scandinavian journal of clinical and laboratory investigation
- Publication Type :
- Academic Journal
- Accession number :
- 27768851
- Full Text :
- https://doi.org/10.1080/00365513.2016.1238507