Back to Search Start Over

Monitoring of chlorsulfuron in biological fluids and water samples by molecular fluorescence using rhodamine B as fluorophore.

Authors :
Alesso M
Escudero LA
Talio MC
Fernández LP
Source :
Talanta [Talanta] 2016 Nov 01; Vol. 160, pp. 431-436. Date of Electronic Publication: 2016 Jul 26.
Publication Year :
2016

Abstract

A new simple methodology is proposed for chlorsufuron (CS) traces quantification based upon enhancement of rhodamine B (RhB) fluorescent signal. Experimental variables that influence fluorimetric sensitivity have been studied and optimized. The zeroth order regression calibration was linear from 0.866 to 35.800µgL(-1) CS, with a correlation coefficient of 0.99. At optimal experimental conditions, a limit of detection of 0.259µgL(-1) and a limit of quantification of 0.866µgL(-1) were obtained. The method showed good sensitivity and adequate selectivity and was applied to the determination of trace amounts of CS in plasma, serum and water samples with satisfactory results analyzed by ANOVA test. The proposed methodology represents an alternative to traditional chromatographic techniques for CS monitoring in complex samples, using an accessible instrument in control laboratories.<br /> (Copyright © 2016 Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1873-3573
Volume :
160
Database :
MEDLINE
Journal :
Talanta
Publication Type :
Academic Journal
Accession number :
27591634
Full Text :
https://doi.org/10.1016/j.talanta.2016.07.053