A new automatic blood pressure kit auscultates for accurate reading with a smartphone: A diagnostic accuracy study.

Authors :
Wu H
Wang B
Zhu X
Chu G
Zhang Z
Source :
Medicine [Medicine (Baltimore)] 2016 Aug; Vol. 95 (32), pp. e4538.
Publication Year :
2016

Abstract

The widely used oscillometric automated blood pressure (BP) monitor was continuously questioned on its accuracy. A novel BP kit named Accutension which adopted Korotkoff auscultation method was then devised. Accutension worked with a miniature microphone, a pressure sensor, and a smartphone. The BP values were automatically displayed on the smartphone screen through the installed App. Data recorded in the phone could be played back and reconfirmed after measurement. They could also be uploaded and saved to the iCloud. The accuracy and consistency of this novel electronic auscultatory sphygmomanometer was preliminarily verified here. Thirty-two subjects were included and 82 qualified readings were obtained. The mean differences ± SD for systolic and diastolic BP readings between Accutension and mercury sphygmomanometer were 0.87 ± 2.86 and -0.94 ± 2.93 mm Hg. Agreements between Accutension and mercury sphygmomanometer were highly significant for systolic (ICC = 0.993, 95% confidence interval (CI): 0.989-0.995) and diastolic (ICC = 0.987, 95% CI: 0.979-0.991). In conclusion, Accutension worked accurately based on our pilot study data. The difference was acceptable. ICC and Bland-Altman plot charts showed good agreements with manual measurements. Systolic readings of Accutension were slightly higher than those of manual measurement, while diastolic readings were slightly lower. One possible reason was that Accutension captured the first and the last korotkoff sound more sensitively than human ear during manual measurement and avoided sound missing, so that it might be more accurate than traditional mercury sphygmomanometer. By documenting and analyzing of variant tendency of BP values, Accutension helps management of hypertension and therefore contributes to the mobile heath service.

Details

Language :
English
ISSN :
1536-5964
Volume :
95
Issue :
32
Database :
MEDLINE
Journal :
Medicine
Publication Type :
Academic Journal
Accession number :
27512876
Full Text :
https://doi.org/10.1097/MD.0000000000004538