Back to Search Start Over

Comparison of the NIST and BIPM Medium-Energy X-Ray Air-Kerma Measurements.

Authors :
Burns DT
O'Brien M
Lamperti P
Boutillon M
Source :
Journal of research of the National Institute of Standards and Technology [J Res Natl Inst Stand Technol] 2003 Oct 01; Vol. 108 (5), pp. 383-9. Date of Electronic Publication: 2003 Oct 01 (Print Publication: 2003).
Publication Year :
2003

Abstract

The air-kerma standards used for the measurement of medium-energy x rays were compared at the National Institute of Standards and Technology (NIST) and at the Bureau International des Poids et Mesures (BIPM). The comparison involved a series of measurements at the BIPM and the NIST using the air-kerma standards and two NIST reference-class transfer ionization standards. Reference beam qualities in the range from 60 kV to 300 kV were used. The results show the standards to be in agreement within the combined standard uncertainty of the comparison of 0.35 %.

Details

Language :
English
ISSN :
1044-677X
Volume :
108
Issue :
5
Database :
MEDLINE
Journal :
Journal of research of the National Institute of Standards and Technology
Publication Type :
Academic Journal
Accession number :
27413616
Full Text :
https://doi.org/10.6028/jres.108.032