Back to Search Start Over

Evaluation Of freehand high-resolution 3-dimensional ultrasound of the median nerve.

Authors :
Pelz JO
Busch M
Weinreich A
Saur D
Weise D
Source :
Muscle & nerve [Muscle Nerve] 2017 Feb; Vol. 55 (2), pp. 206-212. Date of Electronic Publication: 2016 Nov 28.
Publication Year :
2017

Abstract

Introduction: In this study we evaluated freehand 3D ultrasound (3DUS) of the median nerve in comparison to 2D ultrasound (2DUS) and assessed the influence of tilting the transversal plane on cross-sectional area (CSA) measurement.<br />Methods: Two examiners investigated the median nerves of 22 healthy subjects over a distance of 20 cm using 3DUS. Image quality and CSA were assessed at random points within the virtual 3D volume and compared with 2DUS.<br />Results: Image quality within the virtual 3D volume was good/sufficient/poor in 53.0%/40.2%/6.8% (examiner 1) and 21.6%/69.6%/8.8% (examiner 2), respectively. CSA measurements with 3DUS were smaller than with 2DUS (-12% and -17%; Wilcoxon test, P < 0.001). Interrater agreement for 3DUS and intermethod agreement between 2DUS and 3DUS were moderate. Stepwise tilting of the transversal plane increased CSA significantly.<br />Conclusion: Freehand 3DUS of the median nerve over 20 cm is feasible and may help overcome some of the limitations and pitfalls of 2DUS. Muscle Nerve 55: 206-212, 2017.<br /> (© 2016 Wiley Periodicals, Inc.)

Details

Language :
English
ISSN :
1097-4598
Volume :
55
Issue :
2
Database :
MEDLINE
Journal :
Muscle & nerve
Publication Type :
Academic Journal
Accession number :
27387275
Full Text :
https://doi.org/10.1002/mus.25241