Biomechanics of curved surgical needle bending.

Authors :
Abidin MR
Towler MA
Rodeheaver GT
Thacker JG
Cantrell RW
Edlich RF
Source :
Journal of biomedical materials research [J Biomed Mater Res] 1989 Apr; Vol. 23 (A1 Suppl), pp. 129-43.
Publication Year :
1989

Abstract

Because needle bending resistance is such an important factor in the selection of surgical needles, a standard quantitative test of needle bending has been devised to study the biomechanics of deformation of needles from three different commercial manufacturers. The superior resistance to deformation of needles made by one manufacturer has been related to the specific alloy, ASTM 45500 stainless steel, employed in its production. The needles produced from ASTM stainless steel 45500 have stronger tensile and yield strengths than those of the other alloys, accounting for their greater resistance to bending.

Details

Language :
English
ISSN :
0021-9304
Volume :
23
Issue :
A1 Suppl
Database :
MEDLINE
Journal :
Journal of biomedical materials research
Publication Type :
Academic Journal
Accession number :
2722902
Full Text :
https://doi.org/10.1002/jbm.820231312