Back to Search Start Over

Skin dose saving of the staff in 90Y/177Lu peptide receptor radionuclide therapy with the automatic dose dispenser.

Authors :
Fioroni F
Grassi E
Giorgia C
Sara R
Piccagli V
Filice A
Mostacci D
Versari A
Iori M
Source :
Nuclear medicine communications [Nucl Med Commun] 2016 Oct; Vol. 37 (10), pp. 1046-52.
Publication Year :
2016

Abstract

Objective: When handling Y-labelled and Lu-labelled radiopharmaceuticals, skin exposure is mainly due to β-particles. This study aimed to investigate the equivalent dose saving of the staff when changing from an essentially manual radiolabelling procedure to an automatic dose dispenser (ADD).<br />Materials and Methods: The chemist and physician were asked to wear thermoluminescence dosimeters on their fingertips to evaluate the quantity of Hp(0.07) on the skin. Data collected were divided into two groups: before introducing ADD (no ADD) and after introducing ADD.<br />Results: For the chemist, the mean values (95th percentile) of Hp(0.07) for no ADD and ADD are 0.030 (0.099) and 0.019 (0.076) mSv/GBq, respectively, for Y, and 0.022 (0.037) and 0.007 (0.023) mSv/GBq, respectively, for Lu. The reduction for ADD was significant (t-test with P<0.05) for both isotopes. The relative differences before and after ADD collected for every finger were treated using the Wilcoxon test, proving a significantly higher reduction in extremity dose to each fingertip for Lu than for Y (P<0.05). For the medical staff, the mean values of Hp(0.07) (95th percentile) for no ADD and ADD are 0.021 (0.0762) and 0.0143 (0.0565) mSv/GBq, respectively, for Y, and 0.0011 (0.00196) and 0.0009 (0.00263) mSv/GBq, respectively, for Lu. The t-test provided a P-value less than 0.05 for both isotopes, making the difference between ADD and no ADD significant.<br />Conclusion: ADD positively affects the dose saving of the chemist in handling both isotopes. For the medical staff not directly involved with the introduction of the ADD system, the analysis shows a learning curve of the workers over a 5-year period. Specific devices and procedures allow staff skin dose to be limited.

Details

Language :
English
ISSN :
1473-5628
Volume :
37
Issue :
10
Database :
MEDLINE
Journal :
Nuclear medicine communications
Publication Type :
Academic Journal
Accession number :
27218429
Full Text :
https://doi.org/10.1097/MNM.0000000000000548