Back to Search
Start Over
Sustained safety and performance of the second-generation drug-eluting absorbable metal scaffold in patients with de novo coronary lesions: 12-month clinical results and angiographic findings of the BIOSOLVE-II first-in-man trial.
- Source :
-
European heart journal [Eur Heart J] 2016 Sep 14; Vol. 37 (35), pp. 2701-9. Date of Electronic Publication: 2016 May 17. - Publication Year :
- 2016
-
Abstract
- Aims: Metal absorbable scaffolds constitute a conceptually attractive alternative to polymeric scaffolds. Promising 6-month outcomes of a second-generation drug-eluting absorbable metal scaffold (DREAMS 2G), consisting of an absorbable magnesium scaffold backbone, have been reported. We assessed the 12-month safety and performance of this novel device.<br />Methods and Results: The prospective, international, multi-centre, first-in-man BIOSOLVE-II trial enrolled 123 patients with up to two de novo lesions with a reference diameter between 2.2 and 3.7 mm. All patients were scheduled for angiographic follow-up at 6 months, and-if subjects consented-at 12 months. Dual antiplatelet therapy was recommended for 6 months. Quantitative coronary angiography (QCA) parameters remained stable from 6 to 12 months [paired data of 42 patients: in-segment late lumen loss 0.20 ± 0.21 mm vs. 0.25 ± 0.22 mm, P = 0.117, Δ 0.05 ± 0.21 mm (95% CI: -0.01;0.12); in-scaffold late lumen loss 0.37 ± 0.25 mm vs. 0.39 ± 0.27 mm, P = 0.446, Δ 0.03 ± 0.22 (95% CI: -0.04;0.10), respectively]. Intravascular ultrasound and optical coherence tomography findings corroborated the QCA results. Target lesion failure occurred in four patients (3.4%), consisting of one death of unknown cause, one target-vessel myocardial infarction, and two clinically driven target lesion revascularization. No additional event occurred beyond the 6-month follow-up. During the entire follow-up of 12 months, none of the patients experienced a definite or probable scaffold thrombosis.<br />Conclusion: The novel drug-eluting metal absorbable scaffold DREAMS 2G showed a continuous favourable safety profile up to 12 months and stable angiographic parameters between 6 and 12 months.<br />Clinicaltrialsgov Identifier: NCT01960504.<br /> (© The Author 2016. Published by Oxford University Press on behalf of the European Society of Cardiology.)
Details
- Language :
- English
- ISSN :
- 1522-9645
- Volume :
- 37
- Issue :
- 35
- Database :
- MEDLINE
- Journal :
- European heart journal
- Publication Type :
- Academic Journal
- Accession number :
- 27190094
- Full Text :
- https://doi.org/10.1093/eurheartj/ehw196