Improved setup and positioning accuracy using a three-point customized cushion/mask/bite-block immobilization system for stereotactic reirradiation of head and neck cancer.

Authors :
Wang H
Wang C
Tung S
Dimmitt AW
Wong PF
Edson MA
Garden AS
Rosenthal DI
Fuller CD
Gunn GB
Takiar V
Wang XA
Luo D
Yang JN
Wong J
Phan J
Source :
Journal of applied clinical medical physics [J Appl Clin Med Phys] 2016 May 08; Vol. 17 (3), pp. 180-189. Date of Electronic Publication: 2016 May 08.
Publication Year :
2016

Abstract

The purpose of this study was to investigate the setup and positioning uncertainty of a custom cushion/mask/bite-block (CMB) immobilization system and determine PTV margin for image-guided head and neck stereotactic ablative radiotherapy (HN-SABR). We analyzed 105 treatment sessions among 21 patients treated with HN-SABR for recurrent head and neck cancers using a custom CMB immobilization system. Initial patient setup was performed using the ExacTrac infrared (IR) tracking system and initial setup errors were based on comparison of ExacTrac IR tracking system to corrected online ExacTrac X-rays images registered to treatment plans. Residual setup errors were determined using repeat verification X-ray. The online ExacTrac corrections were compared to cone-beam CT (CBCT) before treatment to assess agreement. Intrafractional positioning errors were determined using prebeam X-rays. The systematic and random errors were analyzed. The initial translational setup errors were -0.8 ± 1.3 mm, -0.8 ± 1.6 mm, and 0.3 ± 1.9 mm in AP, CC, and LR directions, respectively, with a three-dimensional (3D) vector of 2.7 ± 1.4 mm. The initial rotational errors were up to 2.4° if 6D couch is not available. CBCT agreed with ExacTrac X-ray images to within 2 mm and 2.5°. The intrafractional uncertainties were 0.1 ± 0.6 mm, 0.1 ± 0.6 mm, and 0.2 ± 0.5 mm in AP, CC, and LR directions, respectively, and 0.0° ± 0.5°, 0.0° ± 0.6°, and -0.1° ± 0.4° in yaw, roll, and pitch direction, respectively. The translational vector was 0.9 ± 0.6 mm. The calculated PTV margins mPTV(90,95) were within 1.6 mm when using image guidance for online setup correction. The use of image guidance for online setup correction, in combination with our customized CMB device, highly restricted target motion during treatments and provided robust immobilization to ensure minimum dose of 95% to target volume with 2.0 mm PTV margin for HN-SABR.

Details

Language :
English
ISSN :
1526-9914
Volume :
17
Issue :
3
Database :
MEDLINE
Journal :
Journal of applied clinical medical physics
Publication Type :
Academic Journal
Accession number :
27167275
Full Text :
https://doi.org/10.1120/jacmp.v17i3.6038