Back to Search Start Over

Limited Expansion of the New Self-Expandable Transcathether Aortic Valve Prosthesis (CoreValve Evolut R).

Authors :
Serio D
Doss M
Kim WK
Möllmann H
Walther T
Source :
The Annals of thoracic surgery [Ann Thorac Surg] 2016 May; Vol. 101 (5), pp. e137-8.
Publication Year :
2016

Abstract

Transcatheter aortic valve implantation (TAVI) has been established as a therapeutic option in patients with a high procedural risk presenting with severe aortic stenosis. Recent improvements of TAVI technology made it possible to treat degenerated bioprosthesis using the valve-in-valve implantation concept. The self-expanding CoreValve (Medtronic, Minneapolis, MN) prosthesis has recently been redesigned and was introduced into clinical practice. We report a case of a not fully expanded Medtronic CoreValve Evolut R after deploying a 26 mm prosthesis into a degenerated 25 mm Carpentier-Edwards Perimount prosthesis.<br /> (Copyright © 2016 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.)