Mechanical testing and modelling of the Universal 2 implant.

Authors :
Gislason MK
Foster E
Main D
Fusiek G
Niewczas P
Bransby-Zachary M
Nash DH
Source :
Medical engineering & physics [Med Eng Phys] 2016 Jun; Vol. 38 (6), pp. 511-7. Date of Electronic Publication: 2016 Apr 12.
Publication Year :
2016

Abstract

Understanding the load mechanics of orthopaedic implants is important to be able to predict their behaviour in-vivo. Much research, both mechanical and clinical, has been carried out on hip and knee implants, but less has been written about the mechanics of wrist implants. In this paper, the load mechanics of the Universal 2 wrist implant have been measured using two types of measuring techniques, strain gauges and Fibre Bragg Grating measurements to measure strains. The results were compared to a finite element model of the implant. The results showed that the computational results were in good agreement with the experimental results. Better understanding of the load mechanics of wrist implants, using models and experimental results can catalyse the development of future generation implants.<br /> (Copyright © 2016 IPEM. Published by Elsevier Ltd. All rights reserved.)

Details

Language :
English
ISSN :
1873-4030
Volume :
38
Issue :
6
Database :
MEDLINE
Journal :
Medical engineering & physics
Publication Type :
Academic Journal
Accession number :
27083120
Full Text :
https://doi.org/10.1016/j.medengphy.2016.02.008