Back to Search Start Over

Diagnostic Accuracy of the HemoCue Hb 301, STAT-Site MHgb and URIT-12 Point-of-Care Hemoglobin Meters in a Central Laboratory and a Community Based Clinic in Durban, South Africa.

Authors :
Jaggernath M
Naicker R
Madurai S
Brockman MA
Ndung'u T
Gelderblom HC
Source :
PloS one [PLoS One] 2016 Apr 05; Vol. 11 (4), pp. e0152184. Date of Electronic Publication: 2016 Apr 05 (Print Publication: 2016).
Publication Year :
2016

Abstract

In South Africa, various point-of-care hemoglobin meters are used. However, the regulatory framework for approval, implementation and oversight of use of point-of-care hemoglobin meters is suboptimal. We assessed the diagnostic accuracy of the HemoCue Hb 301, STAT-Site MHgb and URIT-12 point-of-care hemoglobin meters, compared to a central laboratory based reference assay, in a central laboratory and a community based clinic in Durban, South Africa. Differences in performance of the point-of-care assays, compared to the reference assay, were more pronounced in the community based clinic. Results were reasonable for the HemoCue Hb 301, but poor for the STAT-Site MHgb and the URIT-12. Poor test performance of point-of-care hemoglobin meters, and inadequate evaluations and oversight in South Africa, leads to suboptimal clinical care and clinical research, and increased costs. There is a need for proper evaluation and quality assurance of point-of-care tests, the results of which should be made widely available to key stakeholders.

Details

Language :
English
ISSN :
1932-6203
Volume :
11
Issue :
4
Database :
MEDLINE
Journal :
PloS one
Publication Type :
Academic Journal
Accession number :
27046200
Full Text :
https://doi.org/10.1371/journal.pone.0152184