Numerically simulated exposure of children and adults to pulsed gradient fields in MRI.
- Source :
-
Journal of magnetic resonance imaging : JMRI [J Magn Reson Imaging] 2016 Nov; Vol. 44 (5), pp. 1360-1367. Date of Electronic Publication: 2016 Apr 04. - Publication Year :
- 2016
-
Abstract
- Purpose: To determine exposure to gradient switching fields of adults and children in a magnetic resonance imaging (MRI) scanner by evaluating internal electric fields within realistic models of adult male, adult female, and child inside transverse and longitudinal gradient coils, and to compare these results with compliance guidelines.<br />Materials and Methods: Patients inside x-, y-, and z-gradient coils were simulated using anatomically realistic models of adult male, adult female, and child. The induced electric fields were computed for 1 kHz sinusoidal current with a magnitude of 1 A in the gradient coils. Rheobase electric fields were then calculated and compared to the International Commission on Non-Ionizing Radiation Protection (ICNIRP) 2004 and International Electrotechnical Commission (IEC) 2010 guidelines. The effect of the human body, coil type, and skin conductivity on the induced electric field was also investigated.<br />Results: The internal electric fields are within the first level controlled operating mode of the guidelines and range from 2.7V m <superscript>-1</superscript> to 4.5V m <superscript>-1</superscript> , except for the adult male inside the y-gradient coil (induced field reaches 5.4V m <superscript>-1</superscript> ).The induced electric field is sensitive to the coil type (electric field in the skin of adult male: 4V m <superscript>-1</superscript> , 4.6V m <superscript>-1</superscript> , and 3.8V m <superscript>-1</superscript> for x-, y-, and z-gradient coils, respectively), the human body model (electric field in the skin inside y-gradient coil: 4.6V m <superscript>-1</superscript> , 4.2V m <superscript>-1</superscript> , and 3V m <superscript>-1</superscript> for adult male, adult female, and child, respectively), and the skin conductivity (electric field 2.35-4.29% higher for 0.1S m <superscript>-1</superscript> skin conductivity compared to 0.2S m <superscript>-1</superscript> ).<br />Conclusion: The y-gradient coil induced the largest fields in the patients. The highest levels of internal electric fields occurred for the adult male model. J. Magn. Reson. Imaging 2016;44:1360-1367.<br /> (© 2016 International Society for Magnetic Resonance in Medicine.)
- Subjects :
- Adolescent
Adult
Child
Child, Preschool
Computer Simulation
Female
Humans
Infant
Infant, Newborn
Male
Radiation Dosage
Radiation Protection methods
Young Adult
Aging physiology
Magnetic Fields
Magnetic Resonance Imaging methods
Models, Biological
Radiation Exposure analysis
Radiation Exposure prevention & control
Whole Body Imaging methods
Subjects
Details
- Language :
- English
- ISSN :
- 1522-2586
- Volume :
- 44
- Issue :
- 5
- Database :
- MEDLINE
- Journal :
- Journal of magnetic resonance imaging : JMRI
- Publication Type :
- Academic Journal
- Accession number :
- 27043243
- Full Text :
- https://doi.org/10.1002/jmri.25257