Evaluation of the setup margins for cone beam computed tomography-guided cranial radiosurgery: A phantom study.
- Source :
-
Medical dosimetry : official journal of the American Association of Medical Dosimetrists [Med Dosim] 2016 Autumn; Vol. 41 (3), pp. 199-204. Date of Electronic Publication: 2016 Mar 16. - Publication Year :
- 2016
-
Abstract
- The aim of this study is to evaluate the setup margins from the clinical target volume (CTV) to planning target volume (PTV) for cranial stereotactic radiosurgery (SRS) treatments guided by cone beam computed tomography (CBCT). We designed an end-to-end (E2E) test using a skull phantom with an embedded 6mm tungsten ball (target). A noncoplanar plan was computed (E2E plan) to irradiate the target. The CBCT-guided positioning of the skull phantom on the linac was performed. Megavoltage portal images were acquired after 15 independent deliveries of the E2E plan. The displacement 2-dimensional (2D) vector between the centers of the square field and the ball target on each portal image was used to quantify the isocenter accuracy. Geometrical margins on each patient׳s direction (left-right or LR, anterior-posterior or AP, superior-inferior or SI) were calculated. Dosimetric validation of the margins was performed in 5 real SRS cases: 3-dimesional (3D) isocenter deviations were mimicked, and changes in CTV dose coverage and organs-at-risk (OARs) dosage were analyzed. The CTV-PTV margins of 1.1mm in LR direction, and 0.7mm in AP and SI directions were derived from the E2E tests. The dosimetric analysis revealed that a 1-mm uniform margin was sufficient to ensure the CTV dose coverage, without compromising the OAR dose tolerances. The effect of isocenter uncertainty has been estimated to be 1mm in our CBCT-guided SRS approach.<br /> (Copyright © 2016 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.)
- Subjects :
- Brain Neoplasms pathology
Humans
Organs at Risk
Radiotherapy Dosage
Radiotherapy Planning, Computer-Assisted adverse effects
Radiotherapy, Image-Guided adverse effects
Tumor Burden
Brain Neoplasms radiotherapy
Cone-Beam Computed Tomography methods
Phantoms, Imaging
Radiosurgery methods
Radiotherapy Planning, Computer-Assisted methods
Radiotherapy, Image-Guided methods
Subjects
Details
- Language :
- English
- ISSN :
- 1873-4022
- Volume :
- 41
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- Medical dosimetry : official journal of the American Association of Medical Dosimetrists
- Publication Type :
- Academic Journal
- Accession number :
- 26994824
- Full Text :
- https://doi.org/10.1016/j.meddos.2015.12.006