Back to Search Start Over

An in-situ infection detection sensor coating for urinary catheters.

Authors :
Milo S
Thet NT
Liu D
Nzakizwanayo J
Jones BV
Jenkins ATA
Source :
Biosensors & bioelectronics [Biosens Bioelectron] 2016 Jul 15; Vol. 81, pp. 166-172. Date of Electronic Publication: 2016 Feb 26.
Publication Year :
2016

Abstract

We describe a novel infection-responsive coating for urinary catheters that provides a clear visual early warning of Proteus mirabilis infection and subsequent blockage. The crystalline biofilms of P. mirabilis can cause serious complications for patients undergoing long-term bladder catheterisation. Healthy urine is around pH 6, bacterial urease increases urine pH leading to the precipitation of calcium and magnesium deposits from the urine, resulting in dense crystalline biofilms on the catheter surface that blocks urine flow. The coating is a dual layered system in which the lower poly(vinyl alcohol) layer contains the self-quenching dye carboxyfluorescein. This is capped by an upper layer of the pH responsive polymer poly(methyl methacrylate-co-methacrylic acid) (Eudragit S100®). Elevation of urinary pH (>pH 7) dissolves the Eudragit layer, releasing the dye to provide a clear visual warning of impending blockage. Evaluation of prototype coatings using a clinically relevant in vitro bladder model system demonstrated that coatings provide up to 12h advanced warning of blockage, and are stable both in the absence of infection, and in the presence of species that do not cause catheter blockage. At the present time, there are no effective methods to control these infections or provide warning of impending catheter blockage.<br /> (Copyright © 2016 The Authors. Published by Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1873-4235
Volume :
81
Database :
MEDLINE
Journal :
Biosensors & bioelectronics
Publication Type :
Academic Journal
Accession number :
26945183
Full Text :
https://doi.org/10.1016/j.bios.2016.02.059