Feasibility of Valve-in-Valve Procedure for Degenerated St. Jude Medical Trifecta Bioprosthesis.

Authors :
Verhoye JP
Harmouche M
Soulami RB
Thebault C
Boulmier D
Leguerrier A
Anselmi A
Source :
The Journal of heart valve disease [J Heart Valve Dis] 2015 Jul; Vol. 24 (4), pp. 484-6.
Publication Year :
2015

Abstract

The valve-in-valve (ViV) procedure is an option for patients with symptomatic structural degeneration of a bioprosthesis and excessive reoperative risk. The risk of coronary obstruction appears to be increased if ViV is performed for certain pericardial prostheses in which the leaflets are mounted outside the stent posts. Herein is described a successful ViV for a degenerated Trifecta aortic bioprosthesis, and the technical considerations for performing a ViV procedure within such types of prosthesis are considered. Emphasis is placed on the importance of preoperative investigations (computed tomography scan-based measurements of coronary ostial height and of sinus of Valsalva diameters), and on the precise deployment of the valve (transapical approach with transesophageal echocardiography control) to minimize the risk of major complications. The presence of a failing Trifecta bioprosthesis should not be considered an absolute contraindication to ViV on the basis of the risk of coronary obstruction.