Back to Search
Start Over
Narrowing of the linewidth of an optical parametric oscillator by an acousto-optic modulator for the realization of mid-IR noise-immune cavity-enhanced optical heterodyne molecular spectrometry down to 10⁻¹⁰ cm⁻¹ Hz⁻¹/².
- Source :
-
Optics express [Opt Express] 2015 Dec 28; Vol. 23 (26), pp. 33641-55. - Publication Year :
- 2015
-
Abstract
- The linewidth of a singly resonant optical parametric oscillator (OPO) has been narrowed with respect to an external cavity by the use of an acousto-optic modulator (AOM). This made possible an improvement of the sensitivity of a previously realized OPO-based noise-immune cavity-enhanced optical heterodyne molecular spectrometry instrument for the 3.2 - 3.9 µm mid-infrared region by one order of magnitude. The resulting system shows a detection sensitivity for methane of 2.4 × 10(-10) cm(-1) Hz(-1∕2) and 1.3 × 10(-10) cm(-1) at 20 s, which allows for detection of both the environmentally important (13)CH(4) and CH(3)D isotopologues in atmospheric samples.
Details
- Language :
- English
- ISSN :
- 1094-4087
- Volume :
- 23
- Issue :
- 26
- Database :
- MEDLINE
- Journal :
- Optics express
- Publication Type :
- Academic Journal
- Accession number :
- 26832028
- Full Text :
- https://doi.org/10.1364/OE.23.033641