Back to Search
Start Over
Development of a Mock Extracorporeal Membrane Oxygenation Circuit to Assess Recirculation.
- Source :
-
ASAIO journal (American Society for Artificial Internal Organs : 1992) [ASAIO J] 2016 Jul-Aug; Vol. 62 (4), pp. 496-7. - Publication Year :
- 2016
-
Abstract
- A limitation of veno-venous extracorporeal membrane oxygenation (VV-ECMO) is the recirculating of newly oxygenated blood back to the VV-ECMO circuit. We compared cannulation strategies for combating recirculation utilizing a mock circulation loop (MCL) with ultrasonic flow probes and pressure transducers ensuring that the MCL was run at physiological hemodynamic parameters. Mean recirculation percentages were lower (ANOVA, F = 14.25; p = 0.0001) with the Wang-Zwische dual lumen cannula (4.00 ± 1.77, n = 7) than both the femoro-jugular (15.23 ± 7.00, n = 8) and the femoro-femoral cannulation configurations (13.49 ± 1.44, n = 8).
Details
- Language :
- English
- ISSN :
- 1538-943X
- Volume :
- 62
- Issue :
- 4
- Database :
- MEDLINE
- Journal :
- ASAIO journal (American Society for Artificial Internal Organs : 1992)
- Publication Type :
- Academic Journal
- Accession number :
- 26809084
- Full Text :
- https://doi.org/10.1097/MAT.0000000000000350