A Digital-Analog Microfluidic Platform for Patient-Centric Multiplexed Biomarker Diagnostics of Ultralow Volume Samples.

Authors :
Piraino F
Volpetti F
Watson C
Maerkl SJ
Source :
ACS nano [ACS Nano] 2016 Jan 26; Vol. 10 (1), pp. 1699-710. Date of Electronic Publication: 2016 Jan 12.
Publication Year :
2016

Abstract

Microfluidic diagnostic devices have the potential to transform the practice of medicine. We engineered a multiplexed digital-analog microfluidic platform for the rapid and highly sensitive detection of 3-4 biomarkers in quadruplicate in 16 independent and isolated microfluidic unit cells requiring only a single 5 μL sample. We comprehensively characterized the platform by performing single enzyme and digital immunoassays, achieving single molecule detection and measured as low as ∼10 fM (330 fg/mL) GFP in buffer and ∼12 fM GFP in human serum. We applied our integrated digital detection mechanism to multiplexed detection of 1pM anti-Ebola IgG in human serum and were able to differentiate three common Ebola strains. To ascertain that the device can be applied in environments beyond clinical point-of-care settings, we developed a low-cost, portable hardware system to control and read out the microfluidic device and detected anti-Ebola IgG in ultralow volume whole blood samples to levels of 100 pM in a multiplexed assay format.

Details

Language :
English
ISSN :
1936-086X
Volume :
10
Issue :
1
Database :
MEDLINE
Journal :
ACS nano
Publication Type :
Academic Journal
Accession number :
26741022
Full Text :
https://doi.org/10.1021/acsnano.5b07939