Back to Search Start Over

Modular tapered titanium stems in revision arthroplasty of the hip: The Risk and Causes of Stem Fracture.

Authors :
Konan S
Garbuz DS
Masri BA
Duncan CP
Source :
The bone & joint journal [Bone Joint J] 2016 Jan; Vol. 98-B (1 Suppl A), pp. 50-3.
Publication Year :
2016

Abstract

Tapered fluted titanium stems are increasingly used for femoral revision arthroplasty. They are available in modular and non-modular forms. Modularity has advantages when the bone loss is severe, the proximal femur is mis shapen or the surgeon is unfamiliar with the implant, but it introduces the risk of fracture of the stem at the junction between it and the proximal body segment. For that reason, and while awaiting intermediate-term results of more recently introduced designs of this junction, non-modularity has attracted attention, at least for straightforward revision cases. We review the risks and causes of fracture of tapered titanium modular revision stems and present an argument in favour of the more selective use of modular designs.<br /> (©2016 The British Editorial Society of Bone & Joint Surgery.)

Details

Language :
English
ISSN :
2049-4408
Volume :
98-B
Issue :
1 Suppl A
Database :
MEDLINE
Journal :
The bone & joint journal
Publication Type :
Academic Journal
Accession number :
26733641
Full Text :