Back to Search Start Over

Determination of sulfur in crude oil using high-resolution continuum source molecular absorption spectrometry of the SnS molecule in a graphite furnace.

Authors :
Cadorim HR
Pereira ÉR
Carasek E
Welz B
de Andrade JB
Source :
Talanta [Talanta] 2016 Jan 01; Vol. 146, pp. 203-8. Date of Electronic Publication: 2015 Jul 31.
Publication Year :
2016

Abstract

An analytical method for the determination of sulfur, as the tin mono-sulfide (SnS) molecule, in crude oil using high-resolution continuum source graphite furnace molecular absorption spectrometry (HR-CS GF MAS) has been developed. The molecular absorbance of the SnS has been measured using the wavelength at 271.624 nm and the crude oil samples were prepared as micro-emulsions due to their high viscosity. Several chemical modifiers (Ir, Pd, Ru, Zr) were tested and palladium was chosen, because it exhibited the best performance. The heating program was optimized by comparing the pyrolysis and vaporization curves obtained for an aqueous sulfur standard and a micro-emulsion of a crude oil certified reference material (CRM). The optimum pyrolysis and vaporization temperatures were found to be 600 and 2000°C, respectively. The limit of detection and the characteristic mass using micro-emulsion analysis of crude oil samples were 5.8 and 13.3 ng S. Accuracy and precision of the method has been evaluated using two crude oil CRM (NIST 2721 and NIST 2722), showing good agreement with the informed or certified values.<br /> (Copyright © 2015 Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1873-3573
Volume :
146
Database :
MEDLINE
Journal :
Talanta
Publication Type :
Academic Journal
Accession number :
26695253
Full Text :
https://doi.org/10.1016/j.talanta.2015.07.088