MONTE CARLO STUDY OF THE CARDIAC ABSORBED DOSE DURING X-RAY EXAMINATION OF AN ADULT PATIENT.

Authors :
Kadri O
Manai K
Alfuraih A
Source :
Radiation protection dosimetry [Radiat Prot Dosimetry] 2016 Dec; Vol. 171 (4), pp. 431-437. Date of Electronic Publication: 2015 Oct 13.
Publication Year :
2016

Abstract

The computational voxel phantom 'High-Definition Reference Korean-Man (HDRK-Man)' was implemented into the Monte Carlo transport toolkit Geant4. The voxel model, adjusted to the Reference Korean Man, is 171 cm in height and 68 kg in weight and composed of ∼30 million voxels whose size is 1.981 × 1.981 × 2.0854 mm <superscript>3</superscript> The Geant4 code is then utilised to compute the dose conversion coefficients (DCCs) expressed in absorbed dose per air kerma free in air for >30 tissues and organs, including almost all organs required in the new recommendation of the ICRP 103, due to a broad parallel beam of monoenergetic photons impinging in antero-postero direction with energy ranging from 10 to 150 keV. The computed DCCs of different organs are found to be in good agreement with data published using other simulation codes. Also, the influence of patient size on DCC values was investigated for a representative body size of the adult Korean patient population. The study was performed using five different sizes covering the range of 0.8-1.2 magnification order of the original HDRK-Man. It focussed on the computation of DCC for the human heart. Moreover, the provided DCCs were used to present an analytical parameterisation for the calculation of the cardiac absorbed dose for any arbitrary X-ray spectrum and for those patient sizes. Thus, the present work can be considered as an enhancement of the continuous studies performed by medical physicist as part of quality control tests and radiation protection dosimetry.<br /> (© The Author 2015. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.)

Details

Language :
English
ISSN :
1742-3406
Volume :
171
Issue :
4
Database :
MEDLINE
Journal :
Radiation protection dosimetry
Publication Type :
Academic Journal
Accession number :
26464528
Full Text :
https://doi.org/10.1093/rpd/ncv429