A possible cause of the variable detectability of macroprolactin by different immunoassay systems.

Authors :
Hattori N
Aisaka K
Shimatsu A
Source :
Clinical chemistry and laboratory medicine [Clin Chem Lab Med] 2016 Apr; Vol. 54 (4), pp. 603-8.
Publication Year :
2016

Abstract

Background: Macroprolactinaemia is a major cause of hyperprolactinaemia. The detectability of macroprolactin varies widely among different immunoassay systems, but the causes are not fully known. This study aimed to identify the factors influencing the detectability of macroprolactin by immunoassay systems.<br />Methods: The study included 1544 patients who visited an obstetric and gynaecological hospital. Macroprolactinaemia was screened using the polyethylene glycol (PEG) method and confirmed using gel filtration chromatography and the protein G method. The prolactin (PRL) values determined by enzyme immunoassay (EIA) were compared with those of a chemiluminescence immunoassay system (Centaur) that is known to cross-react the least with macroprolactin.<br />Results: Macroprolactinaemia was found in 62 of 1544 patients (4.02%) who visited an obstetric and gynaecological hospital. The ratio of EIA-determined total PRL to free PRL in the supernatant after PEG precipitation was significantly elevated in all 62 serum samples with macroprolactin compared to those in 1482 serum samples without macroprolactin. In contrast, the ratio of Centaur-determined total PRL to free PRL was significantly elevated in 32 serum samples (group 1) and was within the normal range in 30 (group 2) of 62 serum samples with macroprolactin. The prevalence of non-IgG-type macroprolactin was significantly higher in group 1 than in group 2. Centaur diagnosed hyperprolactinaemia less frequently than EIA (n=2 vs. 16) in 62 patients with macroprolactinaemia. Those two hyperprolactinaemic patients diagnosed by Centaur had non-IgG-type macroprolactin.<br />Conclusions: Macroprolactinaemia was present in 4% of patients visiting an obstetric and gynaecological hospital. The nature of macroprolactin (IgG-type or non-IgG-type) may partly explain why macroprolactin detectability varies among different immunoassay systems.

Details

Language :
English
ISSN :
1437-4331
Volume :
54
Issue :
4
Database :
MEDLINE
Journal :
Clinical chemistry and laboratory medicine
Publication Type :
Academic Journal
Accession number :
26457779
Full Text :
https://doi.org/10.1515/cclm-2015-0484