Hydrodynamic Performance of the Medtronic CoreValve and the Edwards SAPIEN XT Transcatheter Heart Valve in Surgical Bioprostheses: An In Vitro Valve-in-Valve Model.
- Source :
-
The Annals of thoracic surgery [Ann Thorac Surg] 2016 Jan; Vol. 101 (1), pp. 118-24. Date of Electronic Publication: 2015 Sep 10. - Publication Year :
- 2016
-
Abstract
- Background: Valve-in-valve transcatheter aortic valve replacement (TAVR) is becoming a valuable option with promising clinical results in failed bioprosthetic heart valves. Sizing recommendations are based on size compatibility rather than on broad clinical data, in vitro measurements, or biomechanical evidence. The hemodynamic performance of transcatheter heart valves within degenerated surgical heart valves is unknown.<br />Methods: We evaluated the in vitro hydrodynamic performance of two commercially available transcatheter heart valves (Medtronic CoreValve [Medtronic, Minneapolis, MN] and Edwards SAPIEN XT [Edwards Lifesciences, Irvine, CA]) in two different bioprosthetic aortic valves (Edwards Perimount [Edwards Lifesciences] and St. Jude Trifecta [St. Jude Medical, St. Paul, MN]).<br />Results: Within the Edwards Perimount (23 mm) prosthesis, pressure gradients were higher for the SAPIEN XT compared with the CoreValve (11.2 ± 0.1 mm Hg versus 10.1 ± 0.1 mm Hg, p < 0.01), whereas effective orifice area (1.99 ± 0.01 cm(2) versus 1.80 ± 0.01 cm(2), p < 0.01) and total paravalvular leakage (9.0% ± 1.0% versus 5.4% ± 1.3%, p < 0.01) were increased when using the CoreValve. Similarly, measurements in the St. Jude Trifecta revealed higher transvalvular pressure gradients (13.0 ± 0.2 mm Hg versus 10.9 ± 0.3 mm Hg, p < 0.01) and lower effective orifice area for the SAPIEN XT compared with the CoreValve. However, total relative regurgitation was higher with SAPIEN XT as compared with the CoreValve in St. Jude Trifecta prostheses (11.2% ± 1.4% versus 8.3% ± 0.9%, p < 0.01).<br />Conclusions: Both transcatheter heart valve prostheses performed well in the described valve-in-valve settings. Hydrodynamic results were in line with the International Organization for Standardization standards for all configurations. The observed differences indicate a necessity for preclinical valve-in-valve tests in addition to clinical long-term data about longevity.<br /> (Copyright © 2016 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1552-6259
- Volume :
- 101
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- The Annals of thoracic surgery
- Publication Type :
- Academic Journal
- Accession number :
- 26363653
- Full Text :
- https://doi.org/10.1016/j.athoracsur.2015.06.047