Back to Search Start Over

Commonly Practiced Quality Control and Quality Assurance Procedures for Gas Chromatography-Mass Spectrometry Analysis in Forensic Urine Drug-Testing Laboratories.

Authors :
Goldberger BA
Huestis MA
Wilkins DG
Source :
Forensic science review [Forensic Sci Rev] 1997 Dec; Vol. 9 (2), pp. 59-80.
Publication Year :
1997

Abstract

Forensic urine drug-testing laboratories operate in a prescribed scientific and administrative manner to ensure accurate test results. All specimens positive by an initial immunoassay test must be confirmed by gas chromatography/mass spectrometry (GC/MS). To provide adequate control and verification of these analytical processes, laboratories must implement appropriate policies and procedures to be used in routine practice. This review describes the following topics regarding GC/MS analyses: method validation, instrument performance, assay calibration, quality control, criteria for designating a positive test result, sample and batch acceptance criteria, and GC/MS data review.<br /> (Copyright © 1997 Central Police University.)

Details

Language :
English
ISSN :
1042-7201
Volume :
9
Issue :
2
Database :
MEDLINE
Journal :
Forensic science review
Publication Type :
Academic Journal
Accession number :
26269941