Evaluation of the non-calibrated pulse contour cardiac output monitor FloTrac/Vigileo against thermodilution in standing horses.
- Source :
-
Veterinary anaesthesia and analgesia [Vet Anaesth Analg] 2016 Mar; Vol. 43 (2), pp. 153-62. Date of Electronic Publication: 2015 Jun 11. - Publication Year :
- 2016
-
Abstract
- Objective: To evaluate the non-calibrated, minimally invasive cardiac output (CO) monitor FloTrac/Vigileo (FloTrac) against thermodilution (TD) CO in standing horses.<br />Study Design: Prospective, experimental trial.<br />Animals: Nine adult horses weighing a median (range) of 535 (470-602) kg.<br />Methods: Catheters were placed in the right atrium, pulmonary artery and carotid artery under local anaesthesia. CO was measured 147 times by TD and FloTrac and indexed to body weight. Changes in CO were achieved with romifidine or xylazine and dobutamine constant rate infusions. Bland-Altman analysis, concordance and polar plot analysis were used to assess agreement and ability to track changes in CO.<br />Results: Mean ± standard deviation COTD of 48 ± 16 mL kg(-1) minute(-1) (range: 19-93 mL kg(-1) minute(-1) ) and mean COF loTrac of 9 ± 3 mL kg(-1) minute(-1) (range: 5-21 mL kg(-1) minute(-1) ) were measured. Low agreement with a large mean bias of 39 mL kg(-1) minute(-1) and wide limits of agreement of 8-70 mL kg(-1) minute(-1) were found. The percentage error of 108% and precision of TD of ± 18% resulted in an estimated precision of FloTrac of ± 106%. Comparison of changes in COF loTrac with changes in COTD gave a concordance rate of 52% in the four-quadrant plot, and a mean polar angle of -11° with radial limits of agreement of ± 61 ° in the polar plot. Mean arterial pressure (MAP) and COF loTrac were positively correlated (r = 0.5, p < 0.0001). No correlation of MAP with COTD was observed.<br />Conclusions and Clinical Relevance: The FloTrac system, originally designed for use in humans, neither measured absolute CO in standing horses accurately nor tracked relative changes in CO measured by TD correctly. The false dependence of COF loTrac on arterial blood pressure further discourages the use of this technique in horses.<br /> (© 2015 Association of Veterinary Anaesthetists and the American College of Veterinary Anesthesia and Analgesia.)
- Subjects :
- Adrenergic alpha-2 Receptor Agonists administration & dosage
Adrenergic beta-1 Receptor Agonists administration & dosage
Anesthetics administration & dosage
Animals
Blood Pressure
Calibration
Dobutamine administration & dosage
Female
Heart Function Tests instrumentation
Imidazoles administration & dosage
Male
Monitoring, Physiologic methods
Thermodilution
Xylazine administration & dosage
Anesthesia veterinary
Cardiac Output
Heart Function Tests veterinary
Horses
Monitoring, Physiologic veterinary
Subjects
Details
- Language :
- English
- ISSN :
- 1467-2995
- Volume :
- 43
- Issue :
- 2
- Database :
- MEDLINE
- Journal :
- Veterinary anaesthesia and analgesia
- Publication Type :
- Academic Journal
- Accession number :
- 26095722
- Full Text :
- https://doi.org/10.1111/vaa.12281