Back to Search Start Over

The High performance of nanocrystalline CVD diamond coated hip joints in wear simulator test.

Authors :
Maru MM
Amaral M
Rodrigues SP
Santos R
Gouvea CP
Archanjo BS
Trommer RM
Oliveira FJ
Silva RF
Achete CA
Source :
Journal of the mechanical behavior of biomedical materials [J Mech Behav Biomed Mater] 2015 Sep; Vol. 49, pp. 175-85. Date of Electronic Publication: 2015 May 16.
Publication Year :
2015

Abstract

The superior biotribological performance of nanocrystalline diamond (NCD) coatings grown by a chemical vapor deposition (CVD) method was already shown to demonstrate high wear resistance in ball on plate experiments under physiological liquid lubrication. However, tests with a close-to-real approach were missing and this constitutes the aim of the present work. Hip joint wear simulator tests were performed with cups and heads made of silicon nitride coated with NCD of ~10 μm in thickness. Five million testing cycles (Mc) were run, which represent nearly five years of hip joint implant activity in a patient. For the wear analysis, gravimetry, profilometry, scanning electron microscopy and Raman spectroscopy techniques were used. After 0.5 Mc of wear test, truncation of the protruded regions of the NCD film happened as a result of a fine-scale abrasive wear mechanism, evolving to extensive plateau regions and highly polished surface condition (Ra<10nm). Such surface modification took place without any catastrophic features as cracking, grain pullouts or delamination of the coatings. A steady state volumetric wear rate of 0.02 mm(3)/Mc, equivalent to a linear wear of 0.27 μm/Mc favorably compares with the best performance reported in the literature for the fourth generation alumina ceramic (0.05 mm(3)/Mc). Also, squeaking, quite common phenomenon in hard-on-hard systems, was absent in the present all-NCD system.<br /> (Copyright © 2015 Elsevier Ltd. All rights reserved.)

Details

Language :
English
ISSN :
1878-0180
Volume :
49
Database :
MEDLINE
Journal :
Journal of the mechanical behavior of biomedical materials
Publication Type :
Academic Journal
Accession number :
26024650
Full Text :
https://doi.org/10.1016/j.jmbbm.2015.05.005